Temperature Calibration

Temperature Calibration

A very important part of the portable tank gauging tape is the temperature reading, and its accuracy. The TANKTECH Portable System uses a 3-wire PRT or Platinum Resistance Thermometer, which is linear and has standard and stable output all the time.

 A Thermal heated water bath is used, the temperature controller of this water bath has precision PID control mode, giving a minimal deviation of ±0.01șC. Operating temperature range -40șC to +120șC. There is a circulating pump with a capacity of 10 l/min at maximum pressure of 145mbar.

The TECHNE TE-10D has a PRT thermometer, and the indicator has been calibrated against a National Standard probe with error of less than 0.01șC.

An external reference thermometer is used to compare this temperature with the Portable Probe reading. This is the high accuracy Palm Thermometer Probe, with Traceable Accuracy to ±0.05șC or better. It has resolution to 0.01șC or F, and plugs into the Palm without boxes or Modules.

Temperature readings were taken progressively at increments of 5șC from the cold temperature to the highest point of the cargo in the tank for Chemical and Product carriers, without any adjustments done.

          The Palm Thermometer

The readings were recorded with the error readings at each individual point. After it reaches the maximum point, and stabalised for 5 minutes, adjustments are done on the zero and gain of the Analog amplifier inside the probe to achieve a minimum offset reading for the high and low points. For consistent deviations, the zero resistor is adjusted to correct this constant deviation, and for a widening or narrowing deviation, the gain or span resistor is adjusted.

The TFC is a complete digital system on a split architecture, it has two micro-controllers, one at the sensor probe, and one at the display panel. They communicate with each other through the tape conductor bus, and the analogue part of the temperature measurement is inside the probe. Therefore it is only possible to calibrate the probe without the display to get the correct temperature reading.

It is not possible to calibrate the sensor probe without a liquid thermal bath. All calibrations done has a certificate of calibration to go along with the Service Report.

Temperature calibration is a tedious chore, and time consuming. Therefore most Calibration companies charges high price for such task.

Tijit Pte Ltd invested into buying the calibration equipment with the best performance and best control with PID (Proportional, Integral and Derivative) Actions.

 

Digital is our name, Portable is our game