email:
tijit@singnet.com.sg |
| |
 |
Victor 03 Thermo-resistance calibrator |
|
Accuracy : +/- 0.05%
Standard Accessories:
Soft pouch, Test leads, crocodile clamps,
Operating Manual in English.
|
 |
TECHNICAL DATA (Output Function)
|
Output |
Range |
Output range |
Resolution |
Accuracy |
Remarks |
|
Resistance |
400 Ohm |
0.0 to 400.0 Ohm |
0.1 Ohm |
+/- 0.05% Setting value +/- 0.2 Ohm |
1mA excite current Note 1-2 |
|
RTD |
Pt 100 |
-200.0 to 850.0oC |
0.1oC |
+/-0.05% Setting value +/- 0.6oC |
+/-1mA Adopt the Pt100-385 |
|
Note 1-2 |
|
Cu 50 |
-50.0 to 150.0oC |
0.1oC |
1mA excite current Note 1-2 |
|
Measurement
|
|
Input |
Range |
Input Range |
Resolution |
Accuracy |
Remarks |
|
Resistance |
400 Ohm |
0.0 to 400.0 Ohm |
0.1 Ohm |
+/- 0.05% Setting value +/- 0.2 Ohm |
Measuring current 1mA |
|
RTD |
Pt 100 |
-200.0 to 850.0oC |
0.1oC |
+/-0.05% Setting value +/- 0.6oC |
1mA excite current use Pt100-385 |
|
Measuring current 1mA Note 1-2 |
|
Cu 50 |
-50.0 to 150.0oC |
0.1oC |
1mA excite current Note 1-2 |
| Power |
9V battery (IEC 6LR619
alkaline |
| Battery life |
Approx. 20 hours working on 10mA condition. |
| Max. allowable voltage |
30V |
| Operation temperature |
0 to 50șC |
| Operation humidity |
<80% RH |
| Storage Temperature |
<-10șC to +55șC |
| Storage Humidity |
<90% RH |
| Dimensions |
200 x 100 x 40 mm (with
holster) |
| Weight |
550 g (with holster) |
| Accessories |
Operation manual, test line CF-36 (probe with
crocodile clip), soft
pouch |
| Safety standard |
Conforms to IEC 1010 |
|