Loop Calibrator

VICTOR 05 LOOP CALIBRATOR

Third party Certificate

Note:
  1. The max. overload is 1kΩ at 20mA range when the power is higher than 6.8V. The max overload is 700Ω at 20mA range when the power is higher than 5.8 ~ 6.8V.
  2. Power supply range: DC 5~25V.
  3. Temperature coefficient : ±0.005%, range/șC (5șC ~ 18șC, 28șC ~ 40șC).

 Accuracy : ±0.05%

TECHNICAL DATE (Measuring Function)

Input Range Input range Resolution Accuracy Remarks
  Voltage   28V  -0.200 ~28.000V  1mV  +/-0.02% read +2mV   Input 2 M Ohm
  Current   20mA  -1.000~22.000mA  0.001mA  +/-0.02% read +4uA   Input 20M Ohm
 Loop Current   20mA  0.000~22.000mA  0.001mA  +/-0.02% read +4uA   Supply 24V loop power
Output Function Function
Output Range Output range Resolution Accuracy Remarks
  Current  20mA   0.000~22.000mA  0.001mA  +/-0.05% Set value +4uA  20mA the max overload is 1kOhm
  Analog   transducer  -20mA   0.000~22.000mA  0.001mA  +/-0.05% Set value +4uA  20mA, the max overload is 1kOhm 
   
  Loop power  24V      +/-10%   The max output current is 25mA

General Characteristics

 Power  9V battery (IEC 6LR619 alkaline
 Battery life  Approx. 20 hours working on 10mA condition.
 Max. allowable voltage  30V
 Operation temperature  0 to 50șC
 Operation humidity  <80% RH
 Storage Temperature  <-10șC to +55șC
 Storage Humidity  <90% RH
 Dimensions  200 x 100 x 40 mm (with holster)
 Weight  550 g (with holster)
 Accessories  Operation manual, test line CF-36 (probe with crocodile clip), soft pouch
 Safety standard  Conforms to IEC 1010

 