email:
tijit@singnet.com.sg |
| |
|

|
VICTOR 11 PROCESS CALIBRATOR |

|
|
Feature:
6 digit LCD display, output all electrical signals on
field, panel auto-calibrate, cold junction temperature compensation can be
offset automatically. 2-wire, 3-wire or 4-wire connection is selectable on
resistance analogue output.
Auto power-off function can be set by user.
Accuracy
: ±0.02% |
| TECHNICAL DATA (Output
Function) |
|
Output |
Range |
Output range |
Resolution |
Accuracy |
Remarks |
|
Thermo-couple |
R |
-40 ~1760oC |
1oC |
-40~100oC:1.5oC |
Adopt the ITS-90 temperature standard (Note 5) |
|
100~1760oC:1.2oC |
|
S |
-200.0~1760oC |
1oC |
-20~100oC:1.5oC |
|
100~1760oC:1.2oC |
|
B |
-200.0~1800oC |
1oC |
400~600oC:2.0oC |
|
600~800oC:1.5oC |
|
800~1800oC:1.1oC |
|
E |
400~1000oC |
0.1oC |
-200~100oC:0.6oC |
|
-100~600oC: 0.5oC |
|
600~1000oC:0.4oC |
|
K |
-200~1370oC |
0.1oC |
-200~100oC:0.6oC |
|
-100~400oC:0.5oC |
|
100~1200oC:0.7oC |
|
1200~1370oC:0.9oC |
|
T |
-200~400oC |
0.1oC |
-200~400oC:0.6oC |
|
N |
200~1300oC |
0.1oC |
-200~100oC:1.0oC |
|
-100~900oC:0.7oC |
|
900~1300oC:0.8oC |
|
Voltage |
100mV |
-10.0~100.0mV |
0.001mV |
+/-0.02% reading +/10.01% range |
Max. Output current 1mA |
|
1V |
-0.1~1.10000V |
0.01mV |
|
10V |
-1.0~11.0000V |
0.1mV |
|
Current |
20mA |
0.000~22.000mA |
0.001mA |
+/-0.02% reading +/-0.01% range |
20mA the max. overload |
|
is 1k Ohm (note1) |
|
Resistance |
400 Ohm |
0.00~400.00 Ohm |
0.01 Ohm |
+/-0.02% reading +/-0.02% range |
1mA exciting current |
|
(note 2 - note 3) |
|
Analog TX |
-20mA |
0.000~22.000mA |
0.001mA |
+/-0.02% reading |
20mA the max. overload |
|
+/-0.02% range |
is 1 k Ohm (note 4) |
|
RTD |
PT 100 |
-200~850oC |
0.1oC |
-200~0oC:0.3oC |
Adopt the PT100-385 |
|
0~400oC:0.5oC |
|
400~850oC:0.8oC |
(note 2-3) |
|
Cu50 |
-50~150oC |
0.1oC |
-50~150oC:0.6oC |
|
Loop power 24V |
|
|
|
+/-10% |
Max. output current 25mA |
Note:
- When the power is higher than 6.8V, the max overload is 1KΩ
at the range of 20mA; when the power with in 5.8V to 6.8V. the max overload is
700Ω at the range of 20mA.
- Not including the accessorial lead resistance.
- The range of excitation current : 0.5mA to
2mA, the max output voltage <2V.
- Power supply range : DC 5 to 25V
- Not including the accuracy of in-built
temperature compensator transducer. The range of in-built temperature
compensator tranducer : 0 to 50șC, compensation error
<0.5șC.
- Temperature coefficient : ±0.005%. range/șC (5șC to 18șC,
28~50șC)
|
|