email:
tijit@singnet.com.sg |
| |
|
) |
VICTOR 14 Temperature
Calibrator |
 |
| Feature
6-digits LCD Type, unit and value can be displayed by Arrow key.
Auto-calibrate panel, Cold junction compensation can be offset
automatically. RTD 2, 3 or 4 wire connection is selectable. |
Auto-power off timer can be
set by user. User-friendly system. |
|
Accuracy ±0.02%
|
| TECHNICAL DATA (Measuring
Function) |
|
Input |
Range |
Output range |
Resolution |
Accuracy |
Remarks |
|
Voltage |
400mV |
0.00~450.00mV |
10uV |
+/-0.02% reading |
Input resistance 1GΩ |
|
+/-0.01% range |
|
Resistance |
400 Ohm |
0.00~450.00Ω |
0.01Ω |
+/-0.02% reading |
Measuring current 1mA |
|
+/-0.02% range |
(note 1) |
|
THERMOCOUPLE |
R |
-40 ~1760OC |
1OC |
-40~500OC:1.8OC |
Input resistance 1G Ohm Adopts the ITS-90 temperature
standard (note 3) |
|
S |
-20~1760OC |
1OC |
500~1760OC:1.5OC |
|
B |
400~1800OC |
1OC |
400~800OC:2.2OC |
|
800~1000OC:1.8OC |
|
1000~1800OC:1.4OC |
|
E |
-200~1000OC |
0.1OC |
-200~0OC:0.9OC |
|
0~1000OC:1.5OC |
|
K |
-200~1370OC |
0.1OC |
-200~0OC:1.2OC |
|
0~1370OC:
0.8OC |
|
J |
-200~1200OC |
0.1OC |
-200~0OC:1.0OC |
|
O~1200OC:0.7OC |
|
T |
-200~400OC |
0.1OC |
-200~0OC:
1.0OC |
|
0~400OC:
0.7OC |
|
N |
-200~1300OC |
0.1OC |
-200~0OC:
1.5OC |
|
0~1300OC:
0.9OC |
|
RTD |
PT 100 |
-200~850OC |
0.1OC |
-200~0oC:
0.5oC |
Adopt the Pt100-385 1mA Measuring current 1mA
(note 1) |
|
0~400OC:
0.7OC |
|
400~850oC:
0.7oC |
|
Cu50 |
-50~150oC |
0.1OC |
-50~0oC:0.3oC |
|
0~150oC:
0.7oC |
|
| TECHNICAL DATA (Output Function) |
|
Output |
Range |
Input range |
Resolution |
Accuracy |
Remarks |
|
Voltage |
100mV |
-10.000mV~110.000mV |
10uV |
+/-0.02% reading |
The max. output current 1mA |
|
1V |
-0.10000V~1.10000V |
0.01mV |
+/-0.01% range |
|
Resistance |
400Ω |
0.00~450.00Ω |
0.01Ω |
+/-0.02% reading |
1mA exciting current |
|
+/-0.02% range |
(note 1,2) |
|
THERMO-COUPLE |
R |
-40 ~1760OC |
1OC |
-40~100OC:1.5OC |
Input resistance 1G Ohm Adopts the ITS-90 temperature
standard (note 3) |
|
100~1760ºC:1.2ºC |
|
S |
-20~1760OC |
1OC |
-20~100OC:1.5OC |
|
100~1760ºC:1.2ºC |
|
B |
400~1800OC |
1OC |
400~600OC:2.0OC |
|
600~800OC:1.5OC |
|
800~1800OC:1.1OC |
|
E |
-200~1000OC |
0.1OC |
-200~-100OC:0.6OC |
|
-100~600OC:0.5OC |
|
600~1000ºC:0.4ºC |
|
K |
-200~1370OC |
0.1OC |
-200~-100OC:0.6OC |
|
-100~400OC:
0.5OC |
|
400~1200ºC:0.7ºC |
|
1200~1370ºC:0.9ºC |
|
J |
-200~1200OC |
0.1OC |
-200~-100OC:0.6OC |
|
-100~800OC:0.5OC |
|
800~1200ºC:0.7ºC |
|
T |
-200~400OC |
0.1OC |
-200~400OC:
0.6OC |
|
N |
-200~1300OC |
0.1OC |
-200~-100OC:
1.0OC |
|
-100~900OC:
0.7OC |
|
900~1300ºC: 0.8ºC |
|
RTD |
PT 100 |
-200~850OC |
0.1OC |
-200~0oC: 0.3oC |
Adopt the Pt100-385 1mA Measuring current 1mA (note
1) |
|
0~400OC: 0.5OC |
|
400~850oC:
0.8oC |
|
Cu50 |
-50~150oC |
0.1OC |
-50~150oC:0.6oC |
|
Note:
- Not including the accessorial lead resistances.
- The range of exciting current : 0.5mA ~ 2mA, the max.
output voltage : 2V.
- Not including the accuracy of inner temperature
compensatory transducer, the range of inner temperature compensatory
transducer : -10 ~ 50ºC, compensatory error : <0.5ºC.
- Temperature coefficient : ±0.005%, range/ºC (0ºC ~
18ºC, 28ºC ~ 50ºC)
|
|