email:
tijit@singnet.com.sg |
| |
 |
VICTOR 15 V/mA CALIBRATOR |
 |
Feature:
- Display 6-digit LCD.
- Type, unit and value of measurement and output signal
can be displayed by SHIFT key.
- Panel auto-calibrated.
- Auto power-off function can be set by user.
Accuracy
: ±0.02% |
Technical Data (Output Function)
|
Output |
Range |
Output range |
Res. |
Accuracy |
Remarks |
|
Voltage |
10V |
0.0000~11.0000V |
0.1mV |
±0.02% reading |
The max output current is 10mA |
|
±0.01% range |
|
Current |
20mA |
0.000~22.000mA |
0.001mA |
±0.02% reading |
20mA, the max overload is 1kΩ
(Note 1) |
|
0.01% range |
|
Analog transducer |
-20mA |
0.000~22.000mA |
0.001mA |
±0.02% reading |
20mA, the max overload is 1kΩ
(Note 2) |
|
±0.01% range |
|
Loop power |
24V |
|
|
±10% |
The max output current is 25mA |
Measurement Function
|
Input |
Range |
Input range |
Res. |
Accuracy |
Remarks |
|
Voltage |
10V |
0.000~11.000V |
1mV |
±0.02% reading |
Input resistance 2MΩ |
|
±0.01% range |
|
Current |
20mA |
0.000~22.000mA |
0.001mA |
±0.02% reading |
Input resistance 1MΩ |
|
±0.01% range |
|
Loop current |
20mA |
0.000~22.000mA |
0.001mA |
±0.02% reading |
Supply 24V loop power |
|
±0.01% range |
Note:
- The max overload is 1kΩ at 20mA range
when the power is higher than 6.8V.
The max overload is 700 Ω at 20mA
range when the power is higher than 5.8 to 6V.-
Power supply range : DC 5 to 25V.
-
Temperature coefficient :
±0.005% range/șC (0șC to 18șC, 28șC to 50șC).
|