V98A is a handheld meter to
measure electrical parameters. It has all the features of a full-fledge
multi-meter. What differentiates the V98A from the rest, as a pro. of
the pack is: |
- Large LCD displaying three parameters of an input
signal
- Auto-hold, Display hold, and Peak hold to display
the metering value
- Panel calibration function
- 3 data recording modes: LOG, Manual and COMP.
- USB - IR jack to communicate with computer
- Brilliant white LED back light for visual in dark
|

Certificates of approval:
- EMC 2004/108/EC Electromagnetic
Compatibility.
- LVD 2006/95/EC Low Voltage Directive
EN 61010-1: 2003
|
|
Technical Data (Measurement)
|
Input |
Range |
Measuring range |
Res |
Accuracy |
Remark |
|
DC Voltage |
4V |
-4.000V~4.000V |
1MV |
0.2%+4 |
input impedance 10MΩ |
|
40V |
-40.00V~40.00V |
0.01V |
0.2%+4 |
|
400V |
-400.0V~400.0V |
0.1V |
0.2%+4 |
|
1000V |
-1000V~1000V |
1V |
0.2%+4 |
|
AC Voltage (40Hz~600Hz) (5%~100%range) |
400mV |
0~400.0mV |
0.1mV |
1%+4 |
input impedance: 10MΩ<100pF |
|
4V |
0~4.000V |
1mV |
0.5%+4 |
|
40V |
0~40.00V |
0.01V |
0.5%+4 |
|
400V |
0~400.0V |
0.1V |
0.5%+4 |
|
750V |
0~750V |
1V |
1%+4 |
|
DC mV Voltage |
40mV |
-40.00mV~40.00mV |
0.01mV |
0.2%+4 |
input impedance: 10MΩ |
|
|
400mV |
-400.0mV~400.0mV |
0.1mV |
0.2%+4 |
|
OHM |
400Ω |
0~400.0Ω |
0.1Ω |
0.2%+4 |
Plough voltage 0.4V not including the accuracy of down-lead resistance |
|
4kΩ |
0~4.000kΩ |
1Ω |
0.2%+4 |
|
40kΩ |
0~40.00kΩ |
0.01kΩ |
0.2%+4 |
|
400kΩ |
0~400.0kΩ |
0.1kΩ |
0.2%+4 |
|
4MΩ |
0~4.000MΩ |
1kΩ |
0.2%+4 |
|
40MΩ |
0~40.00MΩ |
0.01MΩ |
1%+4 |
|
DC Current |
4A |
-4.000A~4.000A |
0.001A |
1.5%+3 |
input impedance: 0.01Ω |
|
10A |
-10.00A~10.00A |
0.01A |
1.5%+3 |
|
AC current (40Hz~200Hz) (5%~100% rang) |
4A |
-4.000A~4.000A |
0.001A |
1.5%+3 |
input impedance: 0.01Ω |
|
10A |
-10.00A~10.00A |
0.01A |
1.5%+3 |
|
DCmA |
40mA |
-40.00mA~40.00mA |
0.01mA |
0.2%+4 |
input impedance: 1Ω |
|
400mA |
-400.0mA~400.0mA |
0.1mA |
0.2%+4 |
|
AC current (40Hz~500Hz) (5%~100% rang) |
40mA |
0~40.00mA |
0.01mA |
0.5%+4 |
input impedance: 1Ω |
|
400mA |
0~400.0mA |
0.1mA |
0.5%+4 |
|
DCμA |
400μA |
-400.0μA~400.0μA |
0,1μA |
0.2%+4 |
input impedance : 10Ω |
|
4000μA |
-4000μA~4000μA |
1μA |
0.2%+4 |
|
AC current (40Hz~400Hz) (5%~100% rang) |
400μA |
0~400.0μA |
0.1μA |
0.5%+4 |
input impedance: 100Ω |
|
4000μA |
0~4000μA |
1μA |
0.5%+4 |
|
Frequency |
50Hz |
0~50.0Hz |
0.01Hz |
0.1%+3 |
auto range |
|
500Hz |
0~500.0Hz |
0.1Hz |
0.1%+3 |
|
5kHz |
0~5.000kHz |
1Hz |
0.1%+3 |
|
50kHz |
0~50.00kHz |
0.01kHz |
0.1%+3 |
|
100kHz |
0~100.0kHz |
0.1kHz |
0.1%+3 |
|
Duty cycle |
0.1%~99% |
|
0.10% |
1% |
plough voltage : 1.1V~1.6V Current of short circuit: 0.6mA |
|
Diode test |
1V |
|
0.001V |
10% |
|
Continuity test |
<50ΩBB |
|
0.1Ω |
no specs. |
|
Capacitance |
50nF |
0~50.00nF |
0.01nF |
2%+5 |
Auto range time about 30 s in the range of 100μF |
|
|
500nF |
0~500.0nF |
0.1nF |
2%+5 |
|
|
5μF |
0~5.000μF |
0.001μF |
5%+5 |
|
|
50μF |
0~50.00μF |
0.01μF |
5%+5 |
|
|
100μF |
0~100.0μF |
0.1μF |
5%+5 |
|
Thermocouple |
K |
-200~950ºC |
1ºC |
0.5%+2(<-100ºC) |
ITS-90 thermometric scale excludes cold junction accuracy |
|
0.5%+1(>-100ºC) |
|
RTD |
PT100 |
-200~700ºC |
1ºC |
0.5%+2 |
PT-100-385 scale exclude lead resistance error |
|
|
Power |
4 x 1.5V AAA alkaline battery |
|
Working temperature and humidity |
0~50ºC <10ºC,
0~30ºC in 80% RH |
|
Storage temperature and humidity |
-25 to 60ºC, <70%
RH |
|
Working altitude |
<2000m, cannot under height |
|
Shock and vibration |
random 2g, 5~500Hz testing under 1 meter |
|
Proof cycle |
1 year |
|
Warm-up time |
Boot-strap warm up time : about 10 min |
|
Display |
LCD dual display: 68.0 x 36.3mm |
|
Backlight |
White LED backlight can set timer 0~9000s |
|
Accessory standard |
Instruction manual, CD ROM driver software, carrying case, USB
cable |
|
Dimensions and weight |
205 x 95 x 42 (mm) weight about 500g. |
|
|